Focus Microwaves - Vector Load Pull Test Software
Scalar Load Pull
Setup
The typical scalar load pull setup comprises a signal generator, two RF power sensors, a power meter, some DC bias networks and two fundamental tuners. The input and output passive block comprising of couplers, bias tees are also pre-calibrated, and S-Par are defined in the measurement setup.
Parameters
The parameters measured using the scalar loadpull setup are:
- ΓLoad, Γsource
- All spectral quantities like ACPR, EVM and Harmonic Power levels are measured by connecting a Spectrum Analyzer at the output.
Vector Load Pull
Setup
A vector Loadpull setup can be upgraded to hybrid Active Loadpull and time domain measurements using the relevant hardware.
Parameters
The measured RF parameters include:
- PAE, ACPR
- ΓLoad, ΓIN
In this measurement setup both Power In delivered to the DUT and the PAE are calculated using the captured waveforms. All RF parameters are captured in a single shot which dramatically increases the measurement speed. All harmonic power levels are also measured by the Vector receiver.
Pulsed Load Pull
For high power non-linear devices, the self-heating and memory effects severely affect the performance when excited in the continuous wave mode. Using pulsed stimulus signals, the device can be characterized at the higher peak power levels up to saturation to which the devices will be subjected in their intended use. If the devices are excited in pulsed mode, they can be operated at higher peak power level at a reduced risk of device breakdown and a better control of operating temperature. In many applications Pulsed approach represents a realistic operation for devices, like RADARs.
Another important application of pulsed measurements is pulsed I-V and pulsed S-parameter. This approach has been widely used to extract electro thermal models of different device technology. Pulsed operation when combined with high DC voltage levels can also be very interesting for the validation of nonlinear models which considers thermal and trapping effects and can be used to test large devices that could not be tested at the same power levels under CW conditions due to excessive self-heating.
In Pulsed LP measurements both DC and RF can be pulsed. A pulsed load-pull test bench comprises of pulsed bias tees, a primary DC pulse generator, RF source synchronized to a pulse generator, and digitizing scope. The digitizing scope is used to monitor the Pulsed DC characteristics within the pulse. Focus LP uses two different options for Pulse generators, AU5 or MPIV. All passive components of the test setup (including the programmable tuners) are wideband enough to let the pulsed signal pass without any distortion.
For RF characterization there are two different scenarios
1. Scalar LP
Peak power meters synchronized to primary pulsing instrument are used to monitor the RF power within the pulses.
2. Vector LP
Measurements are made using the Vector Network Analyzer receivers which are synchronized with the primary pulsing instrument.
Setup
Paramters
The measured RF parameters include:
- PAE, ACPR
- ΓLoad, ΓIN
Active Load Pull
Active load pull tuning is, because of limited tuning range of passive tuners, additionally reduced by fixture and probe insertion loss, the only method allowing test engineers to reach up to |Γload|=1 (RDUT=0Ω) to match any DUT at its reference plane. In closed loop active tuning, part of the extracted RF power from the DUT is amplified and fed back into the DUT output port, creating a virtual load. Since the re-injected power can be higher than the original extracted power, Γ can be >1, meaning that insertion loss can be compensated. Using two (or more) synchronized (coherent) sources allows open-loop active (and harmonic) tuning; in fact the loop here is also closed, but, instead at RF, it is closed at IF synchronizing reference.
Setup
Parameters
Modulated
WideBand Impedance Tuning
As communication standards require more and more channel bandwidth the need for wideband tuning is increasing. Wideband impedance tuning is now possible as RAPID’s active loop has 100MHz of instantaneous bandwidth allowing users to perform real time modulated measurements. Many spectrum analysis features are also available, such as ACPR, EVM, CCDF, Spectrum mask for advanced modulation standards like LTE and 802.11a/b/g/n/ac.
Active Hybrid Load Pull
Setup
To increase the reflection factor at the probe tip (DUT) and minimize the power loss we must maximize S21 and minimize S11. Any mismatch loss must be compensated by additional injected power in a hybrid (active/passive) tuner.
Parameters
Hybrid tuning is not a panacea. Whereas it allows high VSWR at DUT reference plane, it still remains a rather complex test system with feedback power amplifiers and, often, a second, synchronized, signal source, plus the requirement for in-situ vector power wave measurement, possible through directional couplers inserted between the DUT and the tuner; this on the other hand reduces the tuning range and increases the need for even higher power amplifiers. Passive pre-matching tuning in hybrid systems reduces the requirement for high power from the feedback amplifiers, but only to some extent: passive tuners are not lossless. Tuner loss increases rapidly with reflection factor and so does the power requirement. The critical quantity in tuner loss calculations is “mismatch loss”.
Mismatch loss is S212/(1-S112)
For high S11 values, as needed to pre-match for enhancing the passive reflection factor with active injection in a hybrid configuration, it happens that any increase in insertion loss S21 (due to cables, adapters etc. between tuner and DUT) is multiplied by a factor M=1/(1-S112).
Typical values of the multiplication factor:
S11=0.9 (VSWR=19:1) -> M=5.3
S11=0.96 (VSWR=50:1) -> M=13
Time-Domain Load Pull
The Vector loadpull setup can be extended to fully calibrated time domain measurements by adding a phase reference. A phase reference is a device that has been pre-characterized to uncover the phase relationships between the fundamental and harmonic frequencies.
Time domain characterization technique is aimed at high power, non-linear devices that allows insight in process characteristics, comparisons to knee and breakdown voltages, and to have the ability to place the device into a known class of operation and view the resultant waveform. The shaping of current and voltage waveforms real time in the measurement cycle by accurately terminating the fundamental and harmonic frequencies enables the designer to choose the correct PA mode of operation and how to achieve it.
Setup
Parameters
First step in time domain measurement characterization is the standard small signal calibration, based on established techniques e.g. TRL which are Ratio measurements which ends up with composite error terms. The absolute calibration involves reference to known standards in terms of power and phase. Absolute power calibration is achieved with a calibrated power meter. For relative phase (time) a pre-calibrated phase reference HPR is used.
The measured RF parameters include:
- PAE, ACPR
- ΓLoad, ΓIN
- Loadlines