FRIWO - FOX6- XM-USB 2nd Gen Medical Power Supply 5V / 1400mA worldwide
-
High efficiency: up to 81.6%
-
Low no load power consumption <95mW
-
Meets DoE efficiency level VI
-
CoC version 5 Tier 2
-
Ultra-low leakage current ≤10μA
-
IEC 60601-1-2, 4th edition ready
-
Wide operating temperature range 0 - 45°C
-
Isolation 2x MOPP
-
Double housing safety: snapped & welded
-
IP42 approved interchangeable plug system as additional accessories
-
IEC 6060-1-11 ready with IPX2 adapter
Required Accessories
For the operation of this device you need a primary adapter. This can be found here: Primary adapters
Product information "FOX6- XM-USB 2nd Gen"
- IEC 60601-1 approved
- ≤ 10μA leakage current
- Protection class 2xMOPP
- DOE 2016 ready
- IP42 approved exchangeable plug system as additional accessories
- Double Safety: Housing snapped & welded.
Additional information
- Gehäuse / Housing: FOX6-X-USB
- Protection Class: II
- Output voltage tolerance: ± 5 %
- MTBF: 200.000h acc. Mil217F (based on calculations at 120Vac/60Hz & 230Vac/50Hz, ambient 25°C and 100% load) (The MTBF are theoretically determined values, which does not guarantee the lifetime of the product or of the electrolytic)
-
Storage temperature
: -40 - 70 °C - Working temperature: 0°C - 45°C °C
- Humidity: 10 - 90 %
- AC Input: FRIWO interchangeable mains plug system: see Primary Adapters
- Input voltage: 100 - 240 V
- Input tolerance: ±10%
- Frequency: 50 / 60 Hz
- DC output: USB
- Unit: FOX6-XM-USB
- Current: 1400 mA
- Output voltage: 5 V
- Standard: IEC/EN/UL60601-1 3.1 Edition
- Family : FOX6-X-USB
- Input current: 160 - 80 mA
- leakage current: ≤10 µA
- Overload protection: Yes
- Short Ciruit protection: Yes
- Operating Altitude: ≤3000 m
- Means of Protection: Input Output 2xMOPP, Input - Ground 1xMOPP
- Dimension: 55x31.5x41 mm
- material: PC-ABS
- Spec.0 cover sheet FW-Type: FW8002.1MUSB/05
- Storage Humidity: 10 - 95 %
- Atmospheric pressure: 70 - 106 kPa
- Inrush current: <35
- Line Regulation: ±1
- Load Regulation: ±4
- Start-up-time: ≤2 sec.
- Hold up Zeit: >5ms(120V), >50ms (230V)
- Approbations: Europe, USA, Australia, Japan
- Housing Flame Class: UL 94 V0
- IP Rotection rating: IP40
- Weight: 68 g
- Condcted and Radiated Emmisions: N55032, EN55011, FCC15, Class B, EN60601-1-2
-
Immunity
: EN55024, EN60601-1-2 - Harmonics: EN61000-3-2
- Flicker noise: EN61000-3-3
- ESD (contact/ air): EN61000-4-2
- Immunity against radiated field: EN61000-4-3
- ETF/ Burst: EN61000-4-4
- Surge: EN61000-4-5
- Immunity against conducted disturbances: EN61000-4-6
- Voltage dips: EN61000-4-11
- Immunity against magnetic field: EN61000-4-8
- Drop test: Standard: IEC 60068-2-32, Surface: steel, Drop height: 1 m, Numbers of Drops (total): 3
- Vibration: Standard: IEC 60068-2-6, Frequency range : 10 – 500 Hz, Cross- over-frequency : 58 – 62 Hz, Displacement amplitude : 0.2mm, Peak acceleration : 3 g, Number of cycles: 10 per axis
- color: black