FRIWO - FOX6-XM 2nd Gen Medical Power Supply 5V / 1400mA worldwide
-
High efficiency: up to 81.6%,
-
low no load power consumption <95mW,
-
meets DoE efficiency level VI,
-
CoC version 5 Tier 2,
-
ultra-low leakage current ≤10μA,
-
IEC 60601-1-2, 4th edition ready,
-
wide operating temperature range 0 - 45°C,
-
isolation 2x MOPP,
-
double housing safety: snapped & welded,
-
IP42 approved interchangeable plug system as additional accessories,
-
IEC 60601-1-11 ready with IPX2 adapter
Required Accessories
For the operation of this device you need a primary adapter and a secondary adapter . This can be found here: Primary adapters and Secondary adapters
Additional information
- Output voltage: 5 V
- Family : FOX6
- Current: 1400 mA
- Gehäuse / Housing: FOX6
- Frequency: 50 / 60
- AC Input: FRIWO interchangeable mains plug system: see Primary Adapters
- Humidity: 10 - 90
- Working temperature: 0°C - 40°C °C
-
Storage temperature
: -40 - 70 °C - DC output: Universal output plug system
- Input tolerance: ±10%
- Input voltage: 100-240 V
- IP rating: II
- Spec.0 cover sheet FW-Type: FW8002.1M/05
- color: black
- Input current: 160-80 mA
- leakage current: ≤10 µA
- Overload protection: Yes
- Over Voltage Protection: Yes
- Short Ciruit protection: Yes
- Operating Altitude: 3000 m
- Dimension: 55x31,5x41 mm
- material: PC-ABS
- Standard: IEC60950-1
- Output voltage tolerance: ±5%
- Weight: 108 g
- Storage Humidity: 10-95 %
- Atmospheric pressure: 70-106 kPa
- Line Regulation: ±1
- Load Regulation: ±4
- Hold up Zeit: >5ms(120V) ; >50ms(230V)
- Housing Flame Class: UL 94 V0
- IP Rotection rating: IP40/(IP42 with special Adapter)
- Condcted and Radiated Emmisions: EN55022, EN55011, FCC15, Class B, EN60601-1-2
-
Immunity
: EN55024, EN60601-1-2 - Harmonics: EN61000-3-2
- Flicker noise: EN61000-3-3
- ESD (contact/ air): EN61000-4-2
- Immunity against radiated field: EN61000-4-4
- Surge: EN61000-4-5
- Immunity against conducted disturbances: EN61000-4-5
- Voltage dips: EN61000-4-11
- Immunity against magnetic field: EN61000-4-8
- Drop test: Standard: IEC60068-2-32, Surface: steel, Drop height: 1m, Numbers of Drops (total): 3
- Vibration: Standard: IEC60068-2-6, Frequency range: 10-500Hz, Cross-over-frequenzy:58-62Hz, Displacenebt amplitude:0.2mm, Peak acceleration: 3g, Number of cycles: 10 per axis
- Means of Protection: Input Output 2x MOPP, Input- Ground 1xMOPP
- MTBF: 200.000h acc. Mil217F (based on calculations at
- Approbations: Europe, USA, Australia, Japan