FRIWO - FOX6-XM Medical Power Supply 5V / 1400mA worldwide
-
High efficiency: up to 81.6%
-
Low no load power consumption <95mW
-
Meets DoE efficiency level VI
-
CoC version 5 ready
-
Ultra-low leakage current ≤10μA
-
IEC 60601-1-2, 4th edition ready
-
Wide operating temperature range 0 - 45°C
-
Isolation 2x MOPP
-
Double housing safety: snapped & welded
-
IP42 approved interchangeable plug system as additional accessories
-
Also available as version with fixed plugs: FOX6-FM
Required Accessories
For the operation of this device you need a primary adapter and a secondary adapter . This can be found here:
Primary adapters and Secondary adapters
Primary adapters and Secondary adapters
Product information "FOX6-XM"
- Efficiency and standby losses in accordance with DoE 2016
- IEC 60601-1 approved
- ≤ 10µA leakage current
- Protection class 2xMOPP
- DOE 2016 ready
- IP42 approved exchangeable plug system as additional accessories
- Double Safety: Housing snapped & welded
Additional information
- Output voltage: 5 V
- Current: 1400 mA
- Gehäuse / Housing: FOX6
- Frequency: 50 / 60
- AC Input: FRIWO interchangeable mains plug system: see Primary Adapters
- Humidity: 10% bis 90%
- Working temperature: 0°C - 45°C
-
Storage temperature
: -40°C - 70°C - DC output: Universal output plug system
- Input tolerance: ±10%
- Input voltage: 100-240 V
- Protection Class: II
- Spec.0 cover sheet FW-Type: FW8002M/05
- color: black
- Input current: 160-80 mA
- leakage current: ≤10 µA
- Overload protection: Yes
- Over Voltage Protection: Yes
- Short Ciruit protection: Yes
- Operating Altitude: 3000 m
- Dimension: 55x31,5x41 mm
- material: PC-ABS
- Length: 1830 mm
- Standard: IEC60601-1
- Output voltage tolerance: ±5%
- Weight: 112 g
- Storage Humidity: 10-95%
- Inrush current: <35 A
- Atmospheric pressure: 70-106 kPa
- Line Regulation: ±1
- Load Regulation: ±4
- Hold up Zeit: >5ms(120V) ; >50ms(230V)
- Housing Flame Class: UL 94 V0
- IP Rotection rating: IP40/(IP42 with special Adapter)
- Condcted and Radiated Emmisions: EN55022, EN55011, FCC15, Class B, EN60601-1-2
-
Immunity
: EN55024, EN60601-1-2 - Harmonics: EN61000-3-2
- Flicker noise: EN61000-3-3
- ESD (contact/ air): EN61000-4-2
- Immunity against radiated field: EN61000-4-3
- ETF/ Burst: EN61000-4-4
- Surge: EN61000-4-5
- Immunity against conducted disturbances: EN61000-4-5
- Voltage dips: EN61000-4-11
- Immunity against magnetic field: EN61000-4-8
- Drop test: Standard: IEC60068-2-32, Surface: steel, Drop height: 1m, Numbers of Drops (total): 3
- Vibration: Standard: IEC60068-2-6, Frequency range: 10-500Hz, Cross-over-frequenzy:58-62Hz, Displacenebt amplitude:0.2mm, Peak acceleration: 3g, Number of cycles: 10 per axis
- Means of Protection: Input Output 2x MOPP, Input- Ground 1xMOPP
-
MTBF: 200.000h acc. Mil217F (based on calculations at
120Vac/60Hz & 230Vac/50Hz, ambient 25°C
and 100% load) (The MTBF are theoretically determined values, which does not guarantee the lifetime of the product or of the electrolytic capacitors) - Approbations: Europe, USA, Australia, Japan
Document
Document
Document
More Product Information
Embedded
Download Line Card
Document