FRIWO - FOX30-XM Medical Power Supply with interchangeable adapter system
-
High efficiency: up to 88 %
-
Low no load power consumption < 100 mW
-
Meets DoE efficiency level VI
-
CoC version 5 ready
-
Ultra-low leakage current ≤ 10 μA
-
IEC 60601-1-2, 4th edition ready
-
Wide operating temperature range 0 - 45°C
-
Isolation 2x MOPP
-
Double housing safety: snapped & welded
-
IP42 approved interchangeable plug system as additional accessories
-
Also available as version with fixed plugs: FOX30-FM
FRIWO’s medical products are designed for the most sensitive applications and for use under the harshest conditions. They must continue to work perfectly despite falls in the emergency room while protecting patients by offering the lowest possible levels of leakage current. FRIWO develops and manufactures power supply units you can trust. In order to ensure maximum patient safety, FRIWO medical technology devices are equipped with its 2 x MOPP (Means of Patient Protection) safety system and feature a minimal leakage current of ≤ 10 μA as well as double-sealed housings.
The power supply will operate at universal input voltage range of 90Vac to 264Vac in a wide temperature range of 0-45°C without derating. In addition to patient safety, efficiency plays a major role: with low standby losses of 80mW and high efficiency rates the medical FOX line complies with the demanding US efficiency standard DOE (Department of Energy) and its European counterpart ErP (Energy related Products).
Another particular highlight of the FOX product family is its interchangeable adapter system with IP42 protection (against drops of water and dampness), which is available as an optional accessory. In practice, this means that the devices can be cleaned using wet cloths – an invaluable advantage in sterile environments. FRIWO’s additional features make its products easier to use, day in, day out.
Required Accessories
Primary adapters and Secondary adapters
Product information "FOX30-XM"
- High efficiency: up to 88%
- Low no load power consumption <100mW
- Meets DoE efficiency level VI
- CoC version 5 ready
- Ultra-low leakage current ≤10μA
- IEC 60601-1-2, 4th edition ready
- Wide operating temperature range 0 - 45°C
- Isolation 2x MOPP
- Double housing safety: snapped & welded
- IP42 approved interchangeable plug system as additional accessories
- IEC 60601-1-11 ready with IPX2 adapter
- 4000m operating altitude
Additional information
- Output voltage: 5 V
- Family : FOX30 medical
- Current: 5000 mA
- Means of Protection: Input Output 2xMOPP, Input - Ground 1xMOPP
- Gehäuse / Housing: FOX30 medical
- Output voltage tolerance: ±5%
- Frequency: 50 / 60
- AC Input: FRIWO interchangeable mains plug system: see Primary Adapters
- Humidity: 10% bis 95%
- Working temperature: 0°C - 45°C
- MTBF: 200.000h acc. Mil217F (based on calculations at 120Vac/60Hz & 230Vac/50Hz, ambient 25°C and 100% load) (The MTBF are theoretically determined values, which does not guarantee the lifetime of the product or of the electrolytic ) h
- Standard: IEC/EN/UL60601-1 3.1 Edition
- DC output: Universal output plug system
- Input voltage: 100-240 V
- Protection Class: II
- Spec.0 cover sheet FW-Type: FW8030M/05
- color: black
- Storage temperature: -40-70°C
- leakage current: ≤10 µA
- Overload protection: Yes
- Over Voltage Protection: Yes
- Short Ciruit protection: Yes
- Operating Altitude: 3.000 m
- Dimension: 90x55x34 mm
- Length: 1830 mm
- material: PC-ABS
- Inrush current: <50
- Storage Humidity: 10-95%
- Atmospheric pressure: 70-106 kPa
- Line Regulation: ±1
- Load Regulation: ±4
- Hold up Zeit: >10ms(120V), >50ms (230V)
- Approbations: Europe, USA, Australia, Japan
- Housing Flame Class: UL 94 V0
- IP Rotection rating: IP40
- Condcted and Radiated Emmisions: EN55022, EN55011, FCC15, Class B, EN60601-1-2
- Immunity: EN55024, EN60601-1-2
- Harmonics: EN61000-3-2
- Flicker noise: EN61000-3-3
- ESD (contact/ air): EN61000-4-2
- Immunity against radiated field: EN61000-4-3
- ETF/ Burst: EN61000-4-4
- Surge: EN61000-4-5
- Immunity against conducted disturbances: EN61000-4-6
- Voltage dips: EN61000-4-11
- Immunity against magnetic field: EN61000-4-8
- Vibration: Standard: IEC 60068-2-6, Frequency range : 10 – 500 Hz, Cross- over-frequency : 58 – 62 Hz, Displacement amplitude : 0.2mm, Peak acceleration : 3 g, Number of cycles: 10 per axis