Teledyne LeCroy - PPE1.2KV 10:1/100:1 200/300 MHz 5M Ohm/50M Ohm Probes
The PPE series of probes are suitable for a wide range of applications where high-voltage measurements must be made safely and accurately. There are five fixed-attenuation probes covering a range from 2 kV to 20 kV, and one switchable probe providing ÷10/÷100 attenuation for voltage inputs up to 1.2 kV.
New technology which utilizes hybrid circuitry (and switch reading for probes with switchable gain/attenuation) minimizes ringing and overshoot to provide a precise response.
All fixed-attenuation, standard probes automatically re-scale any LeCroy 9300C or LC series oscilloscope for the appropriate attenuation factor of the probe.
Main Specifications |
|
Bandwidth | 200MHz and 300 MHz |
Attenuation |
10:1 (at 200MHz) / 100:1 (at 300MHz) |
Probes |
|
Probe Ring | No |
Physical Dimensions |
|
Cable Length | 2m |
More Product Information
-
Up to 65 GHz system bandwidth
-
Up to 120 GBaud detectable baud rate
-
Up to 160 GS/s sample rate
-
Real-time acquisition for testing of coherent modulated optical communications links
-
Built-in dispersion compensation, polarization de-multiplexing, and carrier recovery algorithms
-
Supports DP-QPSK, DP-16QAM, and a wide variety of other PSK and QAM formats
-
Support for custom modulation formats
-
Built-in local oscillator
-
Adaptive calibration – Receiver can be disconnected and reconnected without factory calibration
-
Up to 30 GHz bandwidth and 80 GS/s sample rate
-
The most advanced oscilloscope user interface
-
The industry’s only true hardware 14.1 Gb/s serial trigger
-
Low Jitter Measurement Floor and exceptional timebase stability
-
Comprehensive set of serial data analysis, debug, validation and compliance tools
-
Integrated 50 Ω and 1 MΩ inputs for true connection and probing flexibility
-
Integrated standard and custom measurements and math functions for unrivaled analysis capability
-
Multi-lane serial data eye, jitter and crosstalk analysis
-
Real-time de-embedding, emulation, and equalization
-
HD4096 technology provides 12-bit resolution up to 8 GHz and 20 GS/s
-
Up to 5 Gpts of acquisition memory enables detailed viewing of long events
-
15.6" 1900 x 1080 Full HD capacitive touchscreen
-
New ProBus2 input supports up to 8 GHz bandwidth and direct compatibility with a wide variety of existing ProBus probes - 50Ω and 1MΩ coupling modes support all input types on a single connector
-
MAUI with OneTouch user interface for intuitive and efficient operation
-
Deep toolbox enables and simplifies complex analysis
-
Intuitive navigation to quickly find important features in long waveforms
-
6 kV CAT safety rated
-
Widest differential voltage range available
-
Exceptional common-mode rejection ratio (CMRR) across a broad frequency range
-
1% gain accuracy
-
High offset capability at both high and low attenuation
-
AC or DC coupling
-
6 meter cable
-
ProBus active probe interface with automatic scaling
-
Auto-zero capabilities
-
Wide oscilloscope compatibility
-
True Differential TDR with <3mm Spatial Resolution
-
Small Form Factor and Battery Powered
-
S Parameters up to 15 GHz
-
35 ps Rise Time (T3SP15D) 50 ps (T3SP10D)
-
Up to 50,000 points long memory
-
Emerging Serial Data Standards Testing
Transmission line probes are a special type of passive probe designed for use at very high frequencies. They replace the high impedance probe cable found in a traditional passive probe with a precision transmission line, with a characteristic impedance that matches the oscilloscope input (50 Ohm). This greatly reduces the input capacitance to a fraction of a picofarad, minimizing the loading of high frequency signals. A matching network at the tip increases the DC input resistance. While they have lower DC input resistance than a traditional passive probe (usually 500 Ohm) to 5 kOhm), the input impedance of these probes remains nearly constant over their entire frequency range. A traditional /10 passive probe will have a 10 MOhm) input impedance at DC, however this impedance drops rapidly with frequency, passing below the input impedance of a transmission line probe at less than 100 MHz.
-
Static power analysis - like a dedicated power analyzer
-
Dynamic power analysis – capture more information
-
Complete test coverage – control system, power section, motor
-
Comprehensive mechanical interface
-
Harmonics calculation option
-
Vector display option (THREEPHASEVECTOR)
-
Waveform Transformations option (THREEPHASEdq0)
-
16 Channels with OscilloSYNC™
-
2 kV CAT safety rated
-
Only 1500 Vdc safety rated probe per IEC/EN 61010-031:2015
-
Exceptional common-mode rejection ratio (CMRR) across a broad frequency range
-
1% gain accuracy
-
High offset capability at both high and low attenuation
-
AC or DC coupling
-
ProBus active probe interface with automatic scaling
-
Auto-zero capabilities
-
Wide oscilloscope compatibility
With the ProBus interface, the ZD200 becomes an integral part of the oscilloscope. The probe sensitivity and offset are automatically optimized based on the oscilloscope gain settings, so control is transparent to the user. When used with a LeCroy digital oscilloscope, no external power supply is required.
Teledyne LeCroy
Teledyne LeCroy is a leading provider of oscilloscopes, protocol analyzers and related test and measurement solutions that enable companies across a wide range of industries to design and test electronic devices of all types. Since our founding in 1964, we have focused on creating products that improve productivity by helping engineers resolve design issues faster and more effectively.
Oscilloscopes are tools used by designers and engineers to measure and analyze complex electronic signals in order to develop high-performance systems and to validate electronic designs in order to improve time to market. Protocol analyzers are tools used by designers and engineers to generate and monitor traffic over high speed serial data interfaces, including DDR, PCI Express, Fibre Channel, Serial ATA, SAS, USB and others. Both products are critical in the development of a wide variety of demanding design applications.
We utilize both of these important product areas in combination to tackle the most demanding serial data test applications, offering tailored solutions to help our customers perfect their chips, interfaces, subsystems and products which utilize these important communications standards. With the explosion in the use of serial data communications technologies in semiconductors, between devices on circuit boards and between computers and peripherals, we are aimed at a growing and important market.
Our oscilloscopes offer a powerful combination of large and informative displays combined with advanced waveshape analysis capabilities typically tailored to enhance the productivity of engineers in specific applications areas such as serial data test, disk drive test and automotive bus analysis.
Headquartered in Chestnut Ridge, New York, LeCroy has sales, service and development subsidiaries in the US and throughout Europe and Asia. LeCroy products are employed across a wide variety of industries, including semiconductor, computer, consumer electronics, military/aerospace, automotive/industrial, and telecommunications.
Contact Details
Headquarters Teledyne LeCroy
700 Chestnut Ridge Road, Chestnut Ridge, NY 10977-6499, USA
Phone: 800-553-2769 or 845-425-2000
Phone Support: 1-800-553-2769
Fax Sales: 845-578-5985
Email Sales: contact.corp@teledynelecroy.com
Email Support: support@teledynelecroy.com (Oscilloscopes, Waveform Generators, Signal Integrity)